1-866-4 JJSTECH(1-866-455-7832)

Biosystems - Sperian - Honeywell IQ Force 4-Gas Detector - Combustibles (LEL), Oxygen (O2), Duo-Tox (CO / H2S) - IQ-XWY-B-NA

Item #:
IQ-XWY-B-NA
Brand:
Biosystems
$690.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Biosystems - Sperian - Honeywell IQ Force 4-Gas Detector - Combustibles (LEL), Oxygen (O2), Duo-Tox (CO / H2S) - IQ-XWY-B-NA

Biosystems - Sperian - Honeywell IQ Force 4-Gas Detector - Combustibles (LEL), Oxygen (O2), Duo-Tox (CO / H2S) - IQ-XWY-B-NA

$690.00

Discontinued


IQ-XWY-B-NA IQ Force 4-Gas Detector - Combustibles (LEL), Oxygen (O2), Duo-Tox (CO / H2S) by Biosystems

IQ Force multi-gas detectors are simple to use, customizable and reliable—perfect in a variety of industrial applications. This rugged detector monitors for up to four gases simultaneously including oxygen (O2), combustible gases (%LEL), hydrogen sulfide (H2S) and carbon monoxide (CO).

Protecting workers day after day, IQ Force works as a small scale personal gas detection solution or can be combined with an optional pump, docking station and software to provide a more comprehensive IQ Management System solution. A full solution can take your fleet safety program to a new level of efficiency, safety habits and overall safety culture.

Ask us about how the IQ Force can:

  • Improve safety
  • Reduce training
  • Save you from administrative work
  • Improve productivity and reduce downtime
  • Provide confidence
  • Help you build a safety culture

Super bright red LED:

  • Pulses under alarm events
  • Multiple alert modes: SafetyFlash™ or Security/Confidence Flash, both with settable flash frequencies

Easy status confidence:

Flashing heart icon conveys normal detector operation at a glance

Large, easy to read LCD with tricolor backlight:

  • white (normal operation)
  • yellow (low/warning alarms)
  • red (danger/high alarms)

IQ Force Dock:

  • Enables quick, easy and data management
  • Reduces training requirements
  • Saves time on calibration and bump tests
  • Reduces gas costs over manual bump and calibration

Optional motorized sampling pump:

  • Pump module easily attaches in seconds
  • Internal pressure sensor prevents false block alarms
  • IQ Force runs a full shift in pump mode on one charge
IQ Force Specifications:
Housing
Weight 12.8 oz (363 g) without pump; 16.3 oz (463 g) with pump
Detector Dimensions 4.7 x 3.3 x 2.6 in. / 12 x 8.3 x 6.5 cm
Detector with Pump Dimensions 6.3 x 3.3 x 2.6 in. / 16 x 8.3 x 6.5 cm
Environmental Ratings
Temperature Normal Operating: 0º to 110ºF (-18º to 43ºC)
Storage (ideal): 50º to 80ºF (10º to 26ºC)
Humidity 20% to 90% RH continuous, 0% to 95% RH intermittent
Detector IP65/67
Pump IP65; 50 ft. (15m) maximum tubing length
Approvals UL Class I, Division 1, Groups A B C D, T4
UL Class II, Division 1, Groups F G
ATEX Ex d ia IIC T4
IECEX Ex ia IIC T4 Gb
CSA Class I Div 1, Groups A B C D, T4
CSA C22.2 No. 152
Capabilities
Datalogger 41 hours @ 1 minute intervals (default)
Eventlogger Can store up to 200 events
Durability Detector withstands a drop from 6 ft. (1.8 m) onto concrete
IQ Management System Fully compatible. Supports alarm event email, person and location assignment functions
Misc. Able to be docked with or without pump attached
Alarms
Alarm Types High, Low, TWA, STEL, Custom
Visual Bright red LED on top of unit; yellow LCD backlight for low alarm; red LCD backlight for high alarm
Audible Multi-tone, 95 dB @ 1 ft. (30cm)
Vibrating Vibrating alarm standard
Battery
Type Lithium Ion
Run Time Continuous @ 68ºF (20ºC) without pump: 22 hours; with pump: 14 hours
Continuous @ -4ºF (-20ºC) without pump: 13 hours; with pump: 2½ hours
Recharge time 8 hours
Gas Detection Information
Available Sensors O2, %LEL, CO, H2S, Duo-Tox Dual-Channel (CO/H2S), CO-H (hydrogen resistant CO sensor)
Number of Gas Chambers Three sensor capability and four gas capability when using a Duo-Tox (CO/H2S) sensor
Warranty 2 years (monitor), 1 to 2 years sensors, depending on type

IQ Force Sensor Specifications:
Sensor
Range
Resolution
Combustible Gas (LEL)
0-99% LEL, 0-5% CH4
1% LEL, 0.05% CH4
Oxygen (O2)
0-30% by Volume
0.1%
Carbon Monoxide (CO)
Carbon Monoxide, H2 resistant (CO-H)
0-999 ppm
1 ppm
Hydrogen Sulfide (H2S)
0-100 ppm
0.1 or 1 ppm
Duo-Tox Dual-Channel (CO/H2S)
(Provide specific readouts for CO and H2S)
CO: 0-999 ppm
H2S: 0-100 ppm
1 ppm
0.1 or 1 ppm

Questions & Answers