1-866-4 JJSTECH(1-866-455-7832)

RKI Instruments FP-301 Paper Tape Machine for High Sensitivity - FP-301

Item #:
FP-301
Brand:
RKI Instruments
RKI Instruments FP-301 Paper Tape Machine for High Sensitivity - FP-301
$5,880.00
Qty
Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
RKI Instruments FP-301 Paper Tape Machine for High Sensitivity - FP-301

RKI Instruments FP-301 Paper Tape Machine for High Sensitivity - FP-301

$5,880.00

FP-301 Paper Tape Machine for High Sensitivity by RKI Instruments

RKI’s paper tape monitors utilize highly sensitive colorimetric tapes to achieve interference free detection at low PPB and PPM levels of a wide variety of gases. The FP-301 is equipped with an easily replaceable tape that is highly specific to Arsine AsH3 and Hydrogen Selenide H2Se. This model is equipped with a digital display showing the type of gas and concentration in ppb. The FP-301 will also display remaining tape life as well as an end of tape alarm.

The FP-301 has audible and visual alarms, as well as alarm relay contacts and 4-20mA output. A unique “tape saver mode” extends tape life in applications where high background gas levels may be present.

The FP-301 meets and exceeds OSHA and NIOSH compliance requirements, and does not respond to interfering gases (hydrocarbons, alcohols, solvents, CO, etc.) that may be present in industrial/manufacturing environments

Features:

  • Small and compact
  • Highly sensitive and ideal for low level Arsine and Hydrogen Selenide concentrations
  • Interference-free detection
  • Easy detection tape replacement with plug-in cassette
  • Tape remainder display and tape end warning alarm provided
  • PPB digital display
  • Easily transportable
  • 115 VAC powered
  • Alarm relays
  • 4-20mA output

Applications:

  • Furniture and woodworking
  • Coating and varnishing
  • Engineered wood products
  • Medical / veterinary
  • Painting & corrosion resistant finishes
  • Carpet manufacture & storage
  • Adhesives
  • Heat treatment operations
  • Manufactured homes/trailer/RV inspection

*Please note we can only ship RKI products to customers within the United States

Detection principle Photoelectric photometry method using tape cassette
Detection tape Plug in cassette
Detection range 0 - 200ppb
Accuracy Within ± 10% of indication
Tape operation time 1 month with no background gas, or in tape saver mode. Tape remainder display, tape end warning
Alarm Indication Warning and alarm with audible and visual indications
Alarm relay output Form a and b contacts for fail, warning, and alarm rated 0.5A at 125VAC
Alarm functions Latching (standard) or self reset. Lamp (Caution: Yellow, Warning: Red) and buzzer
Self diagnosis Sensor failure, low flow, tape failure, tape end warning, failure of tape setting, system error
External output signal 4-20mA (300 Ω at maximum load resistance)
Operating temperature & humidity 5~35°C 30~90% RH (non condensing)
Power supply & power 24VDC +/- 10% for panel mounting 100 VAC, 110 VAC, 117 VAC, 220 VAC for desktop
Dimensions & weight Panel mounting: Approx. 6.5(W) x 6.5(H) x 10.3(D) Approx. 14.33lb
Desktop: Approx. 6.5(W) x 7.8(H) x 10.4(D) Approx. 12.125lb

Questions & Answers


There are no questions or answers for this item.