1-866-4 JJSTECH(1-866-455-7832)

Atago Polarimeters - POLAX-2L

Item #:
POLAX-2L
Brand:
Atago U.S.A.
Atago Polarimeters - POLAX-2L
$5,650.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Atago Polarimeters - POLAX-2L

Atago Polarimeters - POLAX-2L

$5,650.00

Discontinued


POLAX-2L Polarimeter by Atago U.S.A.

Polarimeter, which is known as an optical rotation meter, is an instrument to measure optical rotation of "optically active matter" such as saccharin, ascorbic acid and sodium glutamate.

Applications

  • Polarimeter is widely used in the pharmaceutical and chemical industries for quality control. Approximately 60 types of chemical materials for measuring “Angle of Rotation” in the list of P)harmacopeia. Including ascorbic acid , testos-terone, and cocaine.
  • Many optically active materials used in food additives, flavoring and perfumes, the polarimetric analysis is widely applied.
  • International sugar scale (°Z) is used for sugar measurement and polarimeter is essential instruments for this procedure.
  • Polarimeter is widely used for educational purpose in the field of “Physics” and “Applied science” course for understanding polarized light, optical rotation, optical active substances, and optical activity.
Measurement Range Angle of rotation :
-179.95° to +180.00°
International sugar scale
-130.0 °Z to +130.0 °Z
Minimum Indication Angle of rotation 0.05° 
International sugar scale
0.1°Z
Measurement Accuracy Angle of rotation ± 0.10°
International sugar scale
±0.3°Z
Measurement temperature range 5 to 40°C
Light Source LED with interference filter (589nm)
Accessories Observation tubes : 100mm tube and 200mm tube / each
Power Supply AC100 to 240V 50/60Hz
Dimensions & Weight 43×22×30cm, 11.5kg (Main Unit only)
Operation switches  R (+) = Right rotation switch : Rotates the analyzer slowly to the right.
 L (-) = Left rotation switch : Rotates the analyzer slowly to the left.
 temp = Shift / Temp switch : Pressing this switch simultaneously with either (+) switch or the (-) switch will provide a faster rotation (in the shift mode).
* When this switch is pressed for 2 seconds or more, the temperature of the sample chamber is indicated on the display panel (until the switch is released from pressing).
Optional • Quartz control plate 34° : RE-72045
• Quartz control plate 17° : RE-72044
• Quartz control plate 8° : RE-72043
* Calibration Certificate : Contact an ATAGO representative for details.
Parts • Observation tube 100mm long OT-100 (P) : RE-6705
• Observation tube 200mm long OT-200 (P) : RE-6706
• Small volume tube 100mm long (OT-100M) : RE-6708
• Deck glass : RE-6712
• Rubber packing for observation tube : RE-6713
• Ring nut for observation tube : RE-6714
• Rubber packing for small volume tube : RE-6715
• Ring nut for small volume tube : RE-6717
• Midget fuse (1A) : RE-5784

Datasheets & User Manuals

Questions & Answers


There are no questions or answers for this item.