RI-257-XXX RI-257 Stand Alone Monitor (Specify Gas) by RKI Instruments

The model RI-257 series are designed to measure the presence of either Fluoro compound (FC) gas, Hydrofluroro compound (HFC or HCFC) gas representing VERTREL XP (HFC-134a) or IPA in the specified range through the built-in continuous operation pump.

As they are based on NDIR (Non-Dispersive Infrared) method, the sample gas specified can be measured without interference from other gases.

Features

  • Compact & light weight (Approx 3.8 kg)
  • Easy to install
  • Negligible interference from other gases
  • Easy to read digital display

Applications

  • Cleaning processes
  • Cold-storage plants
  • Drying rooms
  • Refrigerators

*Please note we can only ship RKI products to customers within the United States

RI-257 Specifications:
Detection principle NDIR (Non-dispersive infra-red)
Measuring gas & range and alarm point See the detection chart below
Sampling method Sample-drawing by built-in pump
Alarm point Dual alarm (WARNING/ALARM)
Run: Latched mode (non-latched after alarm check)
Alarm lamp: Lamp display (Yellow/Red)
Contact output: a contact (b contact option)
Contact rating: AC125V 0.1mA-0.5A (Resistance load)
Repeatability ± 3% of F.S.
Alarm delay time Within 30 seconds by supplying 1.6 times the density gas of alarm preset point
External output 4-20mA (Current outputting type), Max 300 ohms (resistance load)
Warming up 60 mins after turns on power (but measurement starts in 2 mins after power on)
Power source & power consumption Power: AC 100V, AC 117V or AC 220V ± 10%; 50/60 Hz, max 50 VA
Ambient temp/humidity 0 ~ +40°C, 30 – 90% RH, non-condensing
Mounting method Wall mounting
Dimensions Approx 180 (W) x 355 (H) x 97 (D) mm
Weight Approx 3.8 kg
Painting colour Munsell 2.5 Y 9/2
Standard Accessories Fuse, Dust Filter, Instruction Manual, Test Certificate
Optional Accessories Gas Sampling Bag, MC Filter

Gases and ranges detection chart
Gas detection
Range
Alarm set point
FC-14 (*)
0 – 10000 ppm (50 ppm/digit)
3000 ppm
FC-116 (*)
0 – 5000 ppm (25 ppm/digit)
2000 ppm
FC-C318 (*)
0 – 5000 ppm (25 ppm/digit)
2000 ppm
Flon 23
0 – 10000 ppm (50 ppm/digit)
3000 ppm
Flon 32
0 – 10000 ppm (50 ppm/digit)
3000 ppm
Flon 41
0 – 5.000% vol (0.025% vol/digit)
2.000% vol
Flon 134A (Vertrel XP)
0 – 5000 ppm (25 ppm/digit)
2000 ppm
Flon 225
0 – 5000 ppm (25 ppm/digit)
2000 ppm
IPA (Iso-propyl alcohol)
0 – 10000 ppm (50 ppm/digit)
3000 ppm

Datasheets & User Manuals