1-866-4 JJSTECH(1-866-455-7832)

Function Tests on the Instrument

3.6 Function Tests on the Instrument

Alarm Test

Turn ON the instrument. Verify that:

  • all LCD segments are activated momentarily
  • alarm LEDs flash
  • horn sounds briefly
  • vibrating alarm triggers briefly