GasAlertClipExtreme Replacement Parts and Repair Tips

GasAlert Clip Extreme Replacement Parts

DescriptionOrder Number
Universal Front CoverGA24XT-FC1
H2S Specific Front coverGA24XT-H-FC1
Universal Back CoverGA24XT-BC1
Universal LCD replacement GA24XT-LCD
Sensor ScreenGAXT-SS

GasAlert Clip Extreme Repair Tips

Problem Solution
How does user activate monitor? Hold button down for 5 seconds
How often and how should user perform a self test? Daily, hold button down for five seconds in clean environment
How often and how should user calibrate Oxygen sensor At least once per month, press and hold button for three seconds in clean air
Monitor will not respond to gas Replace sensor screen, if it still does not respond to gas, contact local BW office
Monitor is past activation date, can it still be used? Yes but BW does not guarantee full 24 or 36 months of operation

GasAlert Clip Extreme VE

  • Auto-Zero function for low PPM alarms
    • New event type added to event log - records when user initiates auto-zero
    • Enabled when < 5 ppm for H2S, < 25ppm CO, < 3ppm for SO2
    • EDM and IR Data Link don't support new event definition
    • Fleet Manager II supports event
  • E03 Error code added for too much alarm time
  • Button press disabled during alarm condition
  • O2 Calibration time changed
  • Event log resolution changed
    • Resolution for CO is 1ppm from 5
    • H2S, SO2 1ppm (no change)
    • O2 0.1% (no change)

GasAlert Clip Extreme Improvements

  • High Quality Robotic Assembled H2S sensor
  • Auto zero functionality added to toxic versions
  • HOS: The fewer electronic components in a design the fewer there are to fail
  • Vibrating Motor improved from earlier CLIP VE models
GasAlert Clip Extreme Improvements
  • Lower alarm set points
    • Instrument supports fractional alarm set-points
    • Alarm setting for H2S below 10 ppm can be fractional (eg 2.5 ppm)
    • Fractional alarm set points rounded when transferred to Fleet Manager
      • eg 2.4ppm becomes 2ppm, 2.5ppm becomes 3ppm
  • Self test can run in background and can fail detector automatically if sensor fault is detected
    • Functionality of user