Inlet Keys


Inlet Keys

By default, IntelliDoX is configured to connect one calibration gas cylinder to gas inlet 1. In this configuration, gas inlets 2, 3, and 4 are inactive. To connect more than one calibration gas cylinder to IntelliDoX, you must insert a multi-inlet key to activate gas inlets 1, 2, 3, and 4.

Single-inlet Key

Insert the single-inlet key when only one gas inlet is required to support compliance test operations performed on the module
When the single-inlet key is inserted into a module, gas inlet 1 is active and gas inlets 2, 3 and 4 are inactive.

Gas Cylinder with a Single-Inlet Key

When you connect a gas cylinder to IntelliDoX with a single-inlet key, you will see one connected gas type on the LCD screen:


Multi-inlet Key

Insert the multi-inlet key when more than one gas inlet is required to support compliance test operations performed on the module.
When a multi-inlet key is inserted into a module, gas inlets 1, 2 3 and 4 are active.

Connected Gas Cylinders with a Multi-Inlet Key

When you connect multiple gas cylinders to IntelliDoX with a multi-inlet key, you will see multiple connected gas types on the LCD screen:

Change an Inlet Key

To remove the inlet key, pinch the release tabs together, and then pull the inlet key out.
To insert the inlet key, align the inlet key with the opening and push firmly.

NOTICETo remove the inlet key, pinch the release tabs together, and then pull the inlet key out.To insert the inlet key, align the inlet key with the opening and push firmly.

1. Disconnect the power cord from the IntelliDoX
2. On the inserted inlet key, press the release tabs and pull to remove the inlet key.
3. Insert the new inlet key into the IntelliDoX.
4. Connect the power cord to the IntelliDoX.