To replace the cradle, remove the cradle from the IntelliDoX (unscrew the screws and unplug the connector), and then install the replacement cradle.

1. Disconnect the power cord, network cable, tubing, and inlet filter assemblies from the IntelliDoX.
2. Use a screwdriver to unscrew the two screws from the cradle.

3. Lift the cradle, hold the connector at the end of the cable, and pull to unplug the cable from the circuit board.

Do not pull on the cable to unplug the cable. You can cause damage to the cable. Hold the end connector.

4. Remove the cradle from the IntelliDoX.
5. Get the replacement cradle.
6. Plug the replacement cradle's connector cable into the circuit board.
7. Slide the replacement cradle's tabs under the IntelliDoX top assembly to place the cradle into the IntelliDoX.

8. Use a screwdriver to fasten the two screws on the replacement cradle.