Upgrade the Firmware

Upgrade the Firmware

Upgrade the Firmware via IR Link using the Fleet Manager II Software.

Before you begin.

  • You should have the IR Connectivity Kit (sold separately) to transfer the data from a computer to the detector
  • Download and save the firmware update file to a PC or network drive. Do not rename the file.
  • Download the BWFleetManager2.exe file and install Fleet Manager II.
  • For more information refer to the Fleet Manager II User Manual.
  1. Turn on the detector.
  2. Start Fleet Manager II application.
    a) Expand Administration from the left pane
    b) Click Login/Logout .
    c) Type in the default password: Admin .
    d) Click OK to continue.
  3. In the left pane select Devices > Configure device via IR link .
    In the Device Selection window:
    a) Select Honeywell BW ™ Ultra.
    b) Click OK .
  4. In the Honeywell BW ™ Ultra configuration window, click Bootloader to select the binary file.
    In the Honeywell BW ™ Ultra Bootloader window, click Choose File
  5. In the Choose Firmware File to Upload window, select the downloaded file, and then click Open .
  6. Connect the Honeywell BW ™ Ultra detector to the computer using the IR Link connector.
  7. Click Send to initiate the file transfer to the gas detector. After the file transfer is complete, the Bootload process will start. During Bootload, the display will go blank and the detector will beep several times.
  8. The Programming Succeeded message is displayed. Press the button to finish the procedure, and then disconnect the detector from the computer.