1-866-4 JJSTECH(1-866-455-7832)

GfG GMA22-MW/2 Versatile Controller, Configured for 2 measuring points, 4-20mA, power supply not included - 2200-142

Item #:
2200-142
Brand:
GfG
$1,535.10
Qty
Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
GfG GMA22-MW/2 Versatile Controller, Configured for 2 measuring points, 4-20mA, power supply not included - 2200-142

GfG GMA22-MW/2 Versatile Controller, Configured for 2 measuring points, 4-20mA, power supply not included - 2200-142

$1,535.10

2200-142 GMA22-MW/2 Versatile Controller, Configured for 2 measuring points, 4-20mA, power supply not included by GfG

Gases are also used (or may be found) in smaller facilities such as combined heat and power plants or cold stores. A reliable gas detection system therefore makes a significant contribution to the safety of people, your equipment and the environment. We developed the GMA22-M specifically for gas measurement in these applications. Up to four transmitters and relay modules can be connected to the controller. A display indicating the measured values, integrated visual and audible alarms, and four internal relays for additional safety measures complete the GMA22-M's extensive features.

Gases need to be monitored in many processes - regardless of the size of the plant. Even in smaller applications, a gas warning system is indispensable. The GMA22-M is the ideal controller for this purpose, if

  • your application uses up to four transmitters
  • you need to monitor toxic and combustible gases or oxygen
  • a central display of measured values is required outside hazardous areas
  • internal or external relays are required for safety measures
  • you want acoustic and visual alarms to be initiated if gas dangers are present

Wide range of connection options

Using its RS-485 interface, the GMA22-M can be connected to up to four transmitters. In addition to the measured values, the controller also receives digital information such as status messages and service notifications. Alternatively, transmitters can also be connected through an analog 4-20 mA or 0.2-1 mA input.

If you are using any of GfG's ACDC capable transmitters, the analog lines can also transfer digital information on the analog lines (only via 4-20 mA). Many GfG transmitters can be calibrated remotely either using ACDC or digital integration. This simplifies service work significantly, allowing it to be performed by a single technician.

Exterior features

The GMA's dimensions (97 x 140 x 50 mm) are exceptionally small, allowing it to be mounted to a wall almost anywhere. The housing is protected against the ingress of dust and splash water according to IP64, making it suitable for use in demanding external conditions.

The 2.2-inch display comes with multicolor backlight. It lights up green in normal operation and red in the event of an alarm. Alarm levels are indicated by two red LEDs. The integrated horn also gives it an additional acoustic alarm. Two further LEDs indicate the operating status and special states in green and yellow. The pushbutton interface below the display allows for intuitive operation and alarm acknowledgement.

Internal and external relays

The GMA22-M has four integrated relays for initiating safety measures. Up to four additional GMA200-RT relay modules can be connected via the RS-485 interface. With 16 floating signal change-over contacts each, the number of relays is then increased to up to 64. The GMA200Config software is used to configure the GMA22-M and the relay modules. Alarms, faults and measured values can be recorded on a microSD card using a featured data logger, if required.

The GMA22-M is available in four different versions:

  • GMA22-MW (230 V AC): 3x change-over contact, 1x normally open contact, power supply integrated
  • GMA22-MW (24 V DC): 4x change-over contact
  • GMA22-MS (230 V AC): 4x normally open contact, power supply integrated
  • GMA22-MS (24 V DC): 4x normally open contact

Each version is optionally available with a digital or analog input signal.

Key features at a glance:

General information:

  • Suitable for any GfG transmitter
  • Monitors one to four measuring points

Features

  • 4 integrated relays for safety measures
  • 4-20 mA (ACDC capable) or RS-485 (Modbus)
  • Color-backlit 2.2" display
  • Four colored status LEDs
  • Visual and audible alarm
  • Compact housing for wall mounting (IP64)

Optional

  • Data logger (microSD card)
  • With or without integrated power supply
  • Different relay versions
  • Relay module for extension
Display & control elements
Status-LEDs: 4 status LEDs for alarms and operating states
Display: 2,2" graphic display
Buttons: 3 buttons
Alarm: Buzzer max.100dB(A) adjustable
Environmental conditions
Mounting: only indoors up to 2000m above sea level
Storage: -25…+60°C | 5…95%r.h. (recommended 0…+30°C | 40…60%r.h.)
Operating conditions GMA22-MW: -20…+50°C | 5…95%r.h. (without internal PSU)
-20…+45°C | 5…95%r.h. (with internal PSU and max. 150mA load at Uout)
-20…+40°C | 5…95%r.h. (with internal PSU and max. 250mA load at Uout)
Operating conditions GMA22-MS: -20…+55°C | 5…95%r.h. (without internal PSU)
-20…+50°C | 5…95%r.h. (with internal PSU and max. 150mA load at Uout)
-20…+45°C | 5…95%r.h. (with internal PSU and max. 250mA load at Uout)
Power supply (PSU)
  GMA22 with internal PSU GMA22 without internal PSU
Operating voltage Ue: 100-240V AC mains voltage 50-60Hz 24V DC (20-30V DC) through stabilized
SELV or PELV power supply
Power consumption: max. 25VA (with transmitters)
max. 20W (with transmitters)
max. 6VA (without transmitters)
max. 5W (without transmitters)
Fuses: F1: T315mA changeable (for TRM)
F2: 250mA not changeable (for GMA internal)
F1: T630mA changeable (for TRM)
Transmitter connections
  GMA22 with internal PSU GMA22 without internal PSU
Supply output Uout: 24V DC ±5% of internal power supply
max. 300mA
24V DC (20-30V DC) of external power supply
max. 600mA
Analog signal input IIN 4-20mA or 0,2-1mA (4-20mA with ACDC capability)
Tolerance*: ±0,3%MR@4-20mA or ±1,2%MR@0,2-1mA (MR=measuring range)
(load approx. 50..100Ω, Imax=70mA permanent / 500mA temporary)
Digital signals RS485 bus: RS485; Half-Duplex; max. 38400 Baud
Measurement processing
Update time: 1s  
Setting times for RS485: Rise time t50 <2s or t90 <2sec Decay time t50 <2s or t10 <2sec
for 4…20mA: Rise time t50 <2s or t90 <4sec Decay time t50 <2s or t10 <4sec
for 0,2…1mA: Rise time t50 <6s or t90 <10sec Decay time t50 <6s or t10 <10sec
Standby delay: <40s (can be extended by running-in times of gas measuring transmitters)
RS485 output
RS485 bus: RS485; Half-Duplex; max. 38400 Baud (only for GMA200 relay modules)
Relais outputs
Contacts of the GMA22-MS: 4 relays NO
Contacts of the GMA22-MW: 4 relays SPDT (GMA22 without internal PSU)
3 relays SPDT, 1 relay NO (GMA22 with internal PSU)
Insulation distances GMA22-MS: Double insulation between adjacent relays
Insulation distances GMA22-MW: Basic insulation between adjacent relays
Contact load capacity: 3A/250V AC or 3A/30V DC
Minimum switching current: 10mA
Minimum switching voltage: 5V
Data logger (optional) max. 2GB microSD card with FAT formatting (FAT16)
USB connection Mini USB socket for device configuration with PC
Housing
Protection class: IP64 according to IEC 60529; IK08 according to IEC 62262
Material: Plastic
Dimensions: 97 x 140 x 50 mm (W x H x D)
Weight: 275g
Cable junction
Cable: 3-wire ≥0,75mm2 LiYY, NYM (for GMA supply)
2-4 wires 0,5…1,0…(1,5)mm2 LiYY, LiYCY, Y(St)Y (for transmitters)
Cable glands: 5 pieces M16x1,5 (for cable diameter 4,5-10mm)
Terminal blocks: 0,5…1,0mm2 cross section (1,5mm2 for solid conductor)
Approvals / Tests
Electromagnetic compatibility: DIN EN 50270:2015 (Interference emission: type class I, interference immunity: type class II)
Electrical safety: EN 61010-1:2010 (Pollution degree 2, overvoltage category II for mains supply); (Pollution degree 2, overvoltage category III for relay contacts)
Service life 20 years