1-866-4 JJSTECH(1-866-455-7832)

Hanna Portable pH/ORP Meter with SMART Electrode - HI98150

Item #:
HI98150
Brand:
Hanna Instruments
Hanna Portable pH/ORP Meter with SMART Electrode - HI98150
$650.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Hanna Portable pH/ORP Meter with SMART Electrode - HI98150

Hanna Portable pH/ORP Meter with SMART Electrode - HI98150

$650.00

Discontinued


HI98150 Portable pH/ORP Meter with SMART Electrode by Hanna Instruments

HI 98150 is a portable Calibration Check™ pH meter that utilizes HANNA’s SMART electrode technology.

SMART electrodes incorporate a chip, which stores the calibration data performed with a specific instrument. When the “SMART” electrode is attached to the meter again, it is automatically recognized. This allows the operator to optimize time and efficiency with unsurpassed safety. It also avoids erroneous measurements taken in the event the pH electrode is substituted after calibration.

  • pH Calibration Check™ Electrode Diagnostics
  • Waterproof
  • Up to five point pH calibration
  • Up to seven automatically recognized pH buffers and up to two custom pH buffers
  • Log on demand (500 samples)
  • GLP features
  • Auto endpoint
  • User-selectable “calibration time out”
  • Backlit, multi-level LCD display
  • Tutorial messages on LCD
  • Opto-isolated USB
  • BEPS and % battery life on display; Battery Error Prevention System alerts the user in the event that low battery power could adversely affect readings

Order Information:

HI 98150 is supplied with HI 1618D pH electrode, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 70000 electrode cleaning solution sachet, HI 700661 electrode cleaning solution sachet, batteries, instructions and hard carrying case.

pH Range -4.0 to 20.0 pH; -4.00 to 20.00 pH
ORP Range ±699.9; ±2000 mV
Temperature Range -20.0 to 120.0°C (-4.0 to 248.0°F)
pH Resolution 0.1 pH; 0.01 pH
ORP Resolution 0.1 mV (±699.9 mV); 1 mV (outside)
Temperature Resolution 0.1°C (0.1°F)
pH Accuracy @ 20°C ±0.1 pH; ±0.01 pH
ORP Accuracy @ 20°C ±0.2 mV (±699.9 mV); ±1 mV (±2000 mV)
Temperature Accuracy @ 20°C ±0.2°C (±0.4°F) excluding probe error
pH Calibration up to five point calibration with seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and two custom buffers
Offset Calibration ±1 pH
Slope Calibration from 80 to 110%
Temperature Compensation manual or automatic from –20.0 to 120.0 ºC (–4.0 to 248.0 ºF)
pH Electrode HI 1618D PEI body, pre-amplified pH electrode with “SMART” technology, internal temperature sensor, DIN connector and 1 m (3.3’) cable (included)
Logging log on demand, 500 samples
PC Connectivity opto-isolated USB with optional HI 92000 software
Input Impedance 1012 Ohms
Battery Type / Life 1.5V AAA (3) / approximately 200 hours of continuous use without backlight (50 hours with backlight). User selectable auto-off (5, 10, 20, 60 minutes or can be disabled)
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 185 x 72 x 36 mm (7.3 x 2.8 x 1.4”)
Weight 300 g (10.6 oz.)

Datasheets & User Manuals

Questions & Answers


There are no questions or answers for this item.