1-866-4 JJSTECH(1-866-455-7832)

Hanna Turbidity Meter with Fast Tracker™ Technology, EPA Compliant, 230V - HI98703-02

Item #:
HI98703-02
Brand:
Hanna Instruments
Hanna Turbidity Meter with Fast Tracker™ Technology, EPA Compliant, 230V - HI98703-02
$1,157.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Hanna Turbidity Meter with Fast Tracker™ Technology, EPA Compliant, 230V - HI98703-02

Hanna Turbidity Meter with Fast Tracker™ Technology, EPA Compliant, 230V - HI98703-02

$1,157.00

Discontinued


HI98703-02 Turbidity Meter with Fast Tracker™ Technology, EPA Compliant, 230V by Hanna Instruments

The HI 98703 meets and exceeds the requirements of the USEPA Method 180.1 for wastewater and Standard Method 2130 B for drinking water. The instrument has an EPA compliance reading mode which rounds readings to meet EPA reporting requirements. Users will appreciate the accuracy and sensitivity of this instrument, particularly at very low turbidity levels.

This instrument incorporates complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. The last calibration points, time and date can be checked at the touch of a button.

With it’s logging function, up to 200 measurements along with it’s tagged locations can be stored in internal memory and consulted at any time. Data can be later transferred to a PC via RS232 or USB interface and HANNA HI 92000 software (optional).

  • High accuracy at low ranges (below 0.05 NTU)
  • Two, three or four point calibration
  • USB and RS232 PC connectivity
  • Log up to 200 records
  • GLP Features
  • User friendly, backlit display with guidance codes
  • Battery % on startup
  • Continuous current time on display

Exclusive Fast Tracker™

For advanced field applications, the HI 98703 is equipped with Fast Tracker™—Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements.

Order Information:

HI 98703-01 (115V), HI 98703-02 (230V) and HI 98703-03 (AUS plug) are supplied with sample cuvettes and caps (5), HI 98703-11 calibration cuvettes, HI 93703-58 silicone oil, cuvette cleaning cloth, batteries, AC adapter, instruction manual and rugged carrying case.

Range 0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU
Range Selection automatic
Resolution 0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU
Accuracy @25°C ±2% of reading plus 0.02 NTU
Repeatability ±1% of reading or 0.02 NTU, whichever is greater
Stray Light <0.02 NTU
Light Detector silicon photocell
Light Source tungsten filament lamp
Lamp Life greater than 100,000 readings
Method Ratio Nephelometric Method (90°), ratio of scattered and transmitted light; Adaptation of the USEPA Method 180.1 and Standard Method 2130 B
Measuring Mode normal, average, continuous
Turbidity Standards <0.1, 15, 100 and 750 NTU
Calibration two, three or four-point calibration
Log Memory 200 records
PC Connectivity USB or RS232
Environment up to 50°C (122°F); RH max 95% non-condensing
Power Supply 1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use
Dimensions / Weight 224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) / 512 g (18 oz.)

Datasheets & User Manuals

Questions & Answers


There are no questions or answers for this item.