1-866-4 JJSTECH(1-866-455-7832)

Milwaukee Mi160 pH / ORP / ISE / Temperature Laboratory Bench Meter

Item #:
MW-Mi160-US
Brand:
Milwaukee
Milwaukee Mi160 pH / ORP / ISE / Temperature Laboratory Bench Meter
$706.00
Qty
Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Milwaukee Mi160 pH / ORP / ISE / Temperature Laboratory Bench Meter

Milwaukee Mi160 pH / ORP / ISE / Temperature Laboratory Bench Meter

$706.00

Mi160 pH / ORP / ISE / Temperature Laboratory Bench Meter by Milwaukee

This new pH/ORP/ISE/Temp bench meter is ideal for very accurate and precise measurements for all laboratory needs. It can perform ion-selective measurements directly in ppm, as well as pH, ORP and temperature measurements. pH calibration can also be performed in 3 points selectable between 7 memorized buffers, to provide a very accurate calibration curve even when testing different samples, where very wide differences in pH can be found.

Thanks to the memory it can store up to 50 data sets for each range that can be downloaded to a PC via RS232 or USB. These instruments also have GLP features so, at any time, the user can recall the calibration data.

  • 7 memorized buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45) for pH calibration
  • pH calibration up to 3 points
  • ISE calibration up to 2 points; six standard solutions available: 0.01, 0.1, 1, 10, 100, 1000 ppm
  • Messages on the LCD to make the calibration easy and accurate
  • Relative mV feature
  • GLP feature, to view last calibration data for pH or ISE
Range pH: -2.00 to 16.00 pH
mV: ±699.9 mV / ±2000 mV
ISE: 0.001 to 19999 ppm
Temp: -20.0 to 120.0°C / -4.0 to 248.0°F
Resolution pH: 0.01 pH
mV: 0.1 mV / 1 mV
ISE: 0.001 (0.001 to 9.999) ppm; 0.01 (10.00 to 99.99) ppm; 0.1 (100.0 to 999.9) ppm; 1 (1000 to 19999) ppm
Temp: 0.1°C / 0.1°F
Accuracy (@ 20°C) pH: ±0.01 pH
mV: ±0.2 mV / ±1 mV
ISE: ±0.5% Full Scale
Temp: ±0.4°C / ±0.8°F
Rel mV offset ±2000 mV
pH Calibration 1, 2 or 3 point-calibration, with 7 memorized buffers
ISE Calibration 1 or 2 point calibration, 6 standard solutions available
Temperature Compensation automatic, from -20.0 to 120.0°C / -4.0 to 248.0°F or manual, without temperature probe
pH Electrode MA917B/1 (included)
Temperature Probe MA831R (included)
Logging up to 50 records, LOG on demand or auto-logging
Environment 0 to 50°C / 32 to 122°F; max RH 95%
Input Impedance 1012 Ohm
Power Supply 12 VDC power adapter (included)
Dimensions 230 x 160 x 95 mm
Weight 1.1 kg

Questions & Answers


There are no questions or answers for this item.