Milwaukee Mi190 Dissolved Oxygen / Temperature Laboratory Bench Meter

Item #:
MW-Mi190-US
Brand:
Milwaukee
Milwaukee Mi190 Dissolved Oxygen / Temperature Laboratory Bench Meter
$900.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Milwaukee Mi190 Dissolved Oxygen / Temperature Laboratory Bench Meter

Milwaukee Mi190 Dissolved Oxygen / Temperature Laboratory Bench Meter

$900.00

Discontinued


Mi190 Dissolved Oxygen / Temperature Laboratory Bench Meter by Milwaukee

Ideal for testing Dissolved Oxygen in the pharmaceutical and food Industry, as well as monitoring in water treatment plants. The user can choose to measure D.O. readings in mg/L or % of saturation of O2.

This meter can be used for any type of water, as it allows measurements to compensate for temperature, altitude and salinity factors. The automatic logging interval can be set to perform analysis and store data into the memory.

All logged data can be downloaded to your PC through an RS232 or USB serial port. Memory can store up to 50 samples. Mi190 features an automatic calibration procedure, at 1 or 2 points (at 0 and 100% of O2 saturation). The polarographic probe supplied with the meter (MA840) measures the current generated by the reaction of O2 with Ag.

Mi190 is supplied complete with MA840 DO probe with 3 m cable, 2 spare membranes, MA7041 electrolyte solution (30 mL), 12 VDC power adapter, probe holder and instruction manual.

Range O2: 0.00 to 45.00 mg/L (ppm)
% Saturation O2: 0.0 to 300%
Temp: -5.0 to 55.0°C / 23.0 to 131.0°F
Resolution O2: 0.01 mg/L (ppm)
% Saturation O2: 0.1%
Temp: 0.1°C / 0.1°F
Accuracy O2: ±1.5 Full Scale
% Saturation O2: ±1.5 Full Scale
Temp: ±0.4°C / ±0.8°F
Logging 50 records, LOG on demand or auto-logging
DO Calibration automatic, 1 or 2 point at 0% (MA9070) and 100% (in air)
Temperature Compensation 0.0 to 50.0°C / 32.0 to 122.0°F
Altitude Compensation 0 to 4000 m; resolution 100 m
Salinity Compensation 0 to 40 g/L; resolution 1 g/L
DO Probe MA840 with DIN connector (included)
Temperature Probe Included in DO probe
Calibration 2 points (0.0°C and 50.0°C / 32.0 to 122.0°F)
Logging up to 50 records, LOG on demand or auto-logging
PC interface RS232 / USB Opto-isolated
Power supply 12 VDC power adapter (included)
Environment 0 to 50°C / 32 to 122°F; max RH 100%
Dimensions 230 x 160 x 95 mm
Weight 0.9 Kg

Questions & Answers